Interferometric, modulation-free laser stabilization

Not Accessible

Your library or personal account may give you access

Abstract

We demonstrate novel modulation-free frequency locking of a diode laser, utilizing a simple Sagnac interferometer to create an error signal from saturated-absorption spectroscopy. The interference condition at the output of the Sagnac is strongly affected by the sharp dispersion feature near an atomic resonance. Slight misalignment of the interferometer and subsequent spatially selective, or tilt, detection allows this phase change to be converted into an error signal. Tilt locking has significant advantages over previously described methods, as it requires only a small number of low-cost optical components and a detector. In addition, the system has the potential to be constructed as a plug-and-play fiber-coupled monolithic device to provide submegahertz stability for lasers in the commercial market.

© 2002 Optical Society of America

Full Article  |  PDF Article
More Like This
Free-running frequency stability of mid-infrared quantum cascade lasers

T. L. Myers, R. M. Williams, M. S. Taubman, C. Gmachl, F. Capasso, D. L. Sivco, J. N. Baillargeon, and A. Y. Cho
Opt. Lett. 27(3) 170-172 (2002)

Precise frequency measurements of atomic transitions by use of a Rb-stabilized resonator

Ayan Banerjee, Dipankar Das, and Vasant Natarajan
Opt. Lett. 28(17) 1579-1581 (2003)

Sub-Doppler bandwidth atomic optical filter

L. D. Turner, V. Karaganov, P. J. O. Teubner, and R. E. Scholten
Opt. Lett. 27(7) 500-502 (2002)

Cited By

You do not have subscription access to this journal. Cited by links are available to subscribers only. You may subscribe either as an Optica member, or as an authorized user of your institution.

Contact your librarian or system administrator
or
Login to access Optica Member Subscription

Figures (4)

You do not have subscription access to this journal. Figure files are available to subscribers only. You may subscribe either as an Optica member, or as an authorized user of your institution.

Contact your librarian or system administrator
or
Login to access Optica Member Subscription

Equations (1)

You do not have subscription access to this journal. Equations are available to subscribers only. You may subscribe either as an Optica member, or as an authorized user of your institution.

Contact your librarian or system administrator
or
Login to access Optica Member Subscription

Select as filters


Select Topics Cancel
© Copyright 2024 | Optica Publishing Group. All Rights Reserved